TOP
Menu

Thickness Gauge D-1000

Measuring range up to 10, 25 or 50 mm
Resolution 0.01 / 0.001 mm

Thickness Gauge to determine the material thickness of all kinds of fabrics. With these instruments the most popular checking methods can be applied for textiles, felt, geomaterials, floor coverings, leather, foils, rubber etc.

Thickness according to international standards DIN, EN or ISO

CE mark

Special Features


Precise measuring setup

Rigid housing frame

Thickness gauge D-1000 are working according international standards. They measure with highest accuracy.

The instruments are mainly designed for laboratory use

Different measuring feelers and additional weights are available depending on the application area and test standard; on request different standards can be combined in one instrument

The standard measuring range is 0 – 10 mm; 0 – 25 mm and 25 – 50 mm are optional available

The resolution is 0.01 mm; as an option 0.001 mm is available

Thickness gauge D-1000 with USB interface

The feeler is move up and down using a mechanical by hand-wheel

Measuring depth 95 mm (centre of the stamp)

Standard Features


Thickness gauge D-1000 with LCD display – switchable measuring units mm / inch

Gauge with external power supply via the USB interface

The instrument is easy to use

Calibration report for free, DAkkS inspection certificate optional

Available Models

ModelStandardApplicationFeeler
Ø mm
Feeler
cm²
Test Pressure
D-1000-TDIN EN ISO 5084Textiles50.42200.1 kPa and 1 kPa
D-1000-VDIN EN ISO 9073-2Standard Fleece56.42250.5 kPa
D-1000-GDIN EN ISO 9863-1Geomaterials56.42252 kPa
D-1000-P*DIN EN ISO 534Paper162100 kPa
D-1000-HY*DIN EN ISO 54540Hygienic paper56.42252 kPa
D-1000-LDIN EN ISO 2589Leather100.78549,1 kPa
D-1000-GiRubber35.68105 kPa
D-1000-F2*DIN 53370Foil7.5 0.4420.9 N
Feeler
Calotte
D-1000-F1*DIN 53370FoilR 30 mm0.5 N

Resolution 0.001 mm is standard

We also supply custom designed models, which cover different standards within one instrument. Please inform us, which standards you use and what measuring force and feeler size is required. We will send you our suggestion.
The device is also available as mechanical thickness gauge D-2000 and motorized thickness gauge D-2005, D-2006, D-2010 and D-2020 with automatic material feed.
Here you can find a comparison table.

Specifications

Measuring range:0 – 10 mm, Option 0 – 25 mm or 25 – 50 mm
Resolution:0.01 mm, Option 0.001 mm
Accuracy:±0.02 mm, Option ±0.002 mm
Measuring depth:95 mm (centre of the stamp)
Display:LCD 6 digit, 8.5 mm high
Measuring unit:mm or inch (switchable)
Feeler shape and surface pressure:According international Standards DIN, EN and ISO
Output signal:USB
Power supply:Gauge with external power supply via the USB interface
Temperature range:10 – 45 °C
Air humidity:85 % RH, max.
Dimensions:250 x 180 x 400 mm (LxWxH)
Weight, net (gross):Approx. 12 kg (15 kg)

Specifications subject to change without notice!

Attachment (optional)

Code RHigher resolution 0.001 mm instead of 0.01 mm
Code E25 Higher measuring range 0 – 25 mm
Code E50Higher measuring range 0 – 25 mm and 25 – 50 mm
DKD-Cal Calibration report from a DAkkS authorized laboratory

Accessories (optional)

  1. AC adapter

    AC adapter with battery adapter for external power supply.
    Order number: 35011W


Software (optional)

  1. Software P-DMG

    Software P-DMG für DickenmesserSoftware for data transfer, displaying and analysis of measured data on PC

    • Adjusting menu:
      – Language (German/English)
      – Data storage
      – Charge No.
    • Statistical analysis:
      – Actual value
      – Average value
      – Standard deviation
      – Maximum
      – Minimum
      – Register of all readings
    • Data storage in Excel is possible
      Order number: SW-50

Delivery Includes

Thickness Gauge D-1000 with transport packing case, USB cable, calibration report of manufacturer and operating instructions in English or German

If you want to get a quote or want to place an order, please contact us!


+49 (0)8638 9410-0

+49 (0)8638 4825