Thickness Gauge D-2020
Measuring range 0 - 10 or 25 mm, resolution 0.001 mm
with automatic feed
Thickness gage with highest accuracy and resolution.
Thickness Gauge D-2020
Model Information
Print version
Special Features
- Thickness gauge with graphic display for readings and statistics
- Touchscreen for adjusting e. g.: Number of measurings 1 - 50
Measuring path
Measuring time 1 - 60 sec.
Measuring units mm or inch
Feed 10 - 100 mm
2 different travel speeds - 2 selectable modes: "AUTO" - executing a series of measurements
"MAN" - executing a single measurement - Adjustable, automatic feed by series of measurements (travel: 10 - 100 mm, speed: slow or fast)
- Time saving during series measuring by ideal adjustment to the measuring problem
- Statistical analysis of series of measurement (Number of measurings, average value, standard deviation)
- Interface: USB for data transmission
RS 232 for printer - Separate measuring block, incased against outside external impacts, for highest accuracy and repeatability Separate measuring block, incased against outside external impacts, for highest accuracy and repeatability
- Rigid housing frame
- Additional weights for different surface pressure values
Standard Features
- Easy to use
- Standard measuring range 0 - 10 mm
- Thickness gage with selectable measuring units mm / inch
- Resolution 0.001 mm
- Calibration report for free, DKD inspection certificate optional
Available Models
| Model | Standard | Application | Feeler Ø mm | Feeler cm² | Test Pressure |
| D2020-T | DIN EN ISO 5084 | textiles | 50.42 | 20 | 0.1 kPa and 1 kPa |
| D-2020-V | DIN EN ISO 9073-2 | Standard fleece | 56.42 | 25 | 0.1 kPa and 0.5 kPa |
| D-2020-G1 | DIN EN ISO 964-1 | Geomaterials | 56.42 | 25 | 2 kPa and 20 kPa |
| D-2020-NW | DIN EN ISO 53855 | Non-woven textiles | 56.42 35.68 | 25 10 | 0.5 kPa and 1 kPa 5 kPa |
| D-2020-P | DIN EN ISO 53105 | Paper | 16 | 2 | 100 kPa |
| D-2020-HY | DIN EN ISO 54540 | Hygienic paper | 56.42 | 25 | 2 kPa |
| D-2020-L1 | DIN EN ISO 53326 (out of date) | Leather | 10 | 0.785 | 50 kPa |
| D-2020-L2 | DIN EN ISO 2589 | Leather | 10 | 0.785 | 49,1 kPa |
| D-2020-Gi | Rubber | 35.68 | 10 | 5 kPa | |
| Feeler Calotte | |||||
| D-2020-F | DIN 53370 | Foils | R 30 mm | 0.5 N |
Attachment
Code E Higher measuring range 0... 25 mm
DKD-Cal Calibration report from a DKD authorized laboratory
Software (optional)
![]() | P-DMG Software for data transfer, displaying and analysis of measured data on PC Click picture for more information |
Delivery Includes
Specifications
| Measuring range: | 0 - 10 mm, option 25 mm |
| Resolution: | 0.001 mm |
| Accuracy: | ±0.002 mm |
| Depth of jaw: | 160 mm |
| Measuring system: | Capazitive -> Sylvac |
| Display: | Graphic display Touchscreen |
| Measuring unit: | mm or inch (switchable) |
| Stroke in mm: | Level 1 = 2.0, level 2 = 3.5, level 3 = 5.0, level 4 = 6.5 and level 5 = 10.5 mm (full measuring range) |
| Feed: | 10, 20, 25, 40, 50, 60, 80 and 100 mm |
| Residence time: | 1 - 60 seconds |
| Number of measuring: | Up to 50 per test series |
| Feeler shape and surface pressure: | According international standards DIN, EN and ISO |
| Output signal: | RS 232 for printer and USB for PC |
| Temperature range: | 10 - 45 °C |
| Air humidity: | Max. 85 % r.H |
| Dimensions: | 300 x 300 x 320 mm (LxWxH) |
| Weight, net (gross): | approx. 35 kg (approx. 38 kg) |
Specifications subject to change without notice!